OPERATING INSTRUCTIONS | DUSTHUNTER SB30
8014551/1COY/3-0/
2016-08| SICK
Subject to change without notice
TROUBLESHOOTING 6
6.2 Sender/receiver unit
Malfunctions
Warning and error messages in SOPAS ET
Fig. 31: SOPAS ET menu: DH SB30/Diagnosis/Error messages/Warnings
Current warning or error messages, or earlier messages stored in the error memory, can be
shown by selecting “actual” or “memory” in the “Error selection” or “Warnings selection”
window.
Display of error or warning: With LED symbol
Description of error or warning: In the description field of SOPAS ET
Malfunctions listed below can probably be cleared onsite.
Symptom Possible cause Action
LEDs of the
sender/receiver
are not on
No supply voltage
Connection line not connected correctly
or defective
Defective plug connector
Check plug connectors and lines.
Contact SICK Customer Service.
Message Significance Possible cause Action
Span test Deviation from nominal value
above ±2%
Sudden changes in measur-
ing conditions during determi-
nation of control values
Repeat the function check.
Contact SICK Customer Service.
Overflow con-
stant light
Constant light signal > 3.5 V;
measured values invalid
Extraneous light share too
high
Reduce incidence of extraneous light
(select different fitting location, sun
protection, ...).
Measurement
amplifier
Measurement not possible
Measurement receiver not
connected
Check connection to processor board
and connect plug connector, if
necessary (see “Adapting the sender/
receiver unit to the duct diameter”,
page 34).