4 START-UP AND PARAMETER SETTINGS
4.2.9 Starting measurement mode
Set the measuring system to “Measurement” mode after entering/modifying parameters.
To do so, cancel “Maintenance”: Deselect “Maintenance sensor”.
Fig. 25: SOPAS ET menu: MCU/Maintenance/Maintenance
Standard start-up is now completed.
4.2.10 Setting the Ethernet module parameters
Standard setting: 192.168.0.10
A predefined IP address is set on request.
To change the settings:
▸ Select directory “Configuration / I/O Configuration / Interface Module”.
▸ Set the desired network configuration in the “Expansion module information” field and
c
lick “Reset module”.
Fig. 26: SOPAS ET menu: MCU/Configuration/IO configuration/Interface Module
NOTICE:
The risk of undesired access to the measuring system is inherent when communicating
via Ethernet.
▸ Operate the measuring system only behind suitable protection (e.g., Firewall).
The configuration of interface module Ethernet type 2 is not possible with program
SOPAS ET. A special software with description is supplied with the program.