4 START-UP AND PARAMETER SETTINGS
4.2.6 Factory settings
The steps required to modify these settings are described in the following Sections. For this
purpose, the devices must be connected in SOPAS ET (see “Connection to the device via
USB line”, page 41), the Level 1 password set and the “Maintenance” mode set.
▸ Connect the measuring system to the SOPAS ET program.
▸ Enter the Level 1 password (see “Password for SOPAS ET menus”, page 41).
▸ Set the measuring system to “Maintenance” mode.
Fig. 22: SOPAS ET menu: DHSB30/Maintenance/Maintenance
▸ Switch to directory “Application Parameters” and set the desired parameters.
Fig. 23: SOPAS ET menu: Configuration/Application parameter (example)
Entry field Parameter Remark
Mounting location Name of the
measuring location
Sender/receiver unit assignment to the respective measuring location
Response time (T90) Value in s Response time for the measured variable (see “Response time”, page 11)
Setting range 1 ... 600 s
Limit value Value The limit value relay switches when the value entered is overflown or
underflown.
Relay 3 signals Maintenance The relay switched on when “Maintenance” mode is set.
Function check The relay switches on during a function check procedure.
Maintenance request The relay switches on for “Maintenance” mode.
Interval function
check
Time between two
check cycles
see “Function check”, page 12