START-UP AND PARAMETER SETTINGS 4
Live zero Zero point
(2 or 4 mA)
Select 2 or 4 mA to ensure being able to differentiate between measured
value and switched off device or interrupted current loop.
Range low Lower Measuring
Range Limit
Physical value at live zero
Range high Upper Measuring
Range Limit
Physical value at 20 mA
Output check cycle
results
Inactive Control values (see “Function check”, page 12) are not output on the analog
output.
Activated Control values (see “Function check”, page 12) are output on the analog
output.
cc2 Quadratic Entry of the regression factors determined using gravimetric comparison
measurement during a calibration (see “Calibration for dust concentration
measurement”, page 46)
cc1 Linear
cc0 Absolute
Entry field Parameter Remark