33
OPERATING INSTRUCTIONS | DUSTHUNTER SB30
8014551/1COY/3-0/
2016-08| SICK
Subject to change without notice
ASSEMBLY AND INSTALLATION 3
3.3.4.3 Standard connection (without options)
Fig. 16: Standard connection
NOTE:
Only one RS485 interface may be connected each time (either to plug
connector 1 or to plug connector 3). Otherwise communication malfunctions
can result.
Sender/receiver unit
3
4
5
6
1
2
7
Plug assignment
(View on pin side)
Plug connector 1
Voltage supply, interface
Plug connector 3
Digital inputs,
service
Plug connector 2
Analog output,
relay outputs
Plug assignment
(View on pin side)
Socket assignment
(View on pin side)
3
4
5
6
1
2
7
Pin Assignment
1+24 V DC
2CAN high
3RS485 B
4RS485 A
5CAN low
6-24 V DC
7 Screen
Pin Assignment
1
Maintenance
2
Function check/linearity
measurement
3
RS485 B (Service)
4
RS485A (Service)
5
gnd
Pin Assignment
1
+20 mA
2
-20 mA
3
Relay 1 (operation/failure)
4
Relay 2 (limit value)
5
Relay 3 (maintenance/function
check/maintenance request)
6
Com. (Rel. 1 - 3)
7
Screen
2
1
4
3
5