56 FLOWSIC100 Flare · Addendum to Operating Instructions · 8025365/V1-0/2020-05 · © SICK Engineering GmbH
Installation
9 The self-cutting ring remains in the fixed position on the duct probe.
10 Then tighten the pipe screw fitting hand-tight.
4.8.3 Installing sender/receiver units
1 Make sure the ball valves are closed.
If applicable, close ball valves.
If applicable, remove blind flanges.
2 Position the flange gasket.
3 Position the sender/receiver unit on the ball valve.
Make sure the seal does not move during positioning.
Ensure for cross-duct versions that the Slave sender/receiver unit (→ Fig. 35) is
installed on the nozzle on the downstream side so that the sender/receiver unit points
against the flow direction:
4 Insert the four bolts and screw the sender/receiver unit on the ball valve (torque: see
Operating Instructions FLOWSIC100 Flare §6.7 "Sealing installation instruction").
5 Completely loosen the pipe screw fitting.
6 Open the ball valve.
7 Push the sender/receiver unit into the pipeline.
8 Check the gasket of the pipe screw fitting for damage.
9 Screw in the pipe screw fitting and tighten with a torque of 150 Nm.
10 Push the sender/receiver unit further to the stop.
11 For probe version F1F-P, align the measuring path correctly before the self-cutting ring
fitting is fixed. For cross-duct versions, proceed with the next step.
12 Tighten the self-cutting ring fitting 1.25 turns.
Make sure the markings for the self-cutting ring fitting are next to each other again,
Fig. 32.
13 For cross-duct versions, install the Master sender/receiver unit on the nozzle on the
upstream side so that the sender/receiver unit points in the flow direction.
Observe when aligning the probe version
Align probe version F1F-P correctly before securing the self-cutting ring fitting again:
WARNING: Leakage hazard
Multiple use can damage the gasket of the pipe screw fitting.
Before reuse, i.e. whenever the pipe screw fitting is to be tightened again,
check the gasket of the pipe screw fitting:
Replace the gasket if it has visible deformations, indentations or damage. In
this case contact SICK Service.
Otherwise there is a risk of leakage.
WARNING: Danger due to gas leaks
When gas escapes, close the ball valve again and contact SICK Service.
WARNING: Leakage hazard
Multiple use can damage the gasket of the pipe screw fitting.
Before reuse, i.e. whenever the pipe screw fitting is to be tightened again,
check the gasket of the pipe screw fitting:
Replace the gasket if it has visible deformations, indentations or damage. In
this case contact SICK Service.
Otherwise there a risk of leakage.

Table of Contents

Other manuals for SICK FLOWSIC100 Flare

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK FLOWSIC100 Flare and is the answer not in the manual?

SICK FLOWSIC100 Flare Specifications

General IconGeneral
BrandSICK
ModelFLOWSIC100 Flare
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals