56 FLOWSIC100 Flare · Addendum to Operating Instructions · 8025365/V1-0/2020-05 · © SICK Engineering GmbH
Installation
9 The self-cutting ring remains in the fixed position on the duct probe.
10 Then tighten the pipe screw fitting hand-tight.
4.8.3 Installing sender/receiver units
1 Make sure the ball valves are closed.
– If applicable, close ball valves.
– If applicable, remove blind flanges.
2 Position the flange gasket.
3 Position the sender/receiver unit on the ball valve.
Make sure the seal does not move during positioning.
Ensure for cross-duct versions that the Slave sender/receiver unit (→ Fig. 35) is
installed on the nozzle on the downstream side so that the sender/receiver unit points
against the flow direction:
4 Insert the four bolts and screw the sender/receiver unit on the ball valve (torque: see
Operating Instructions FLOWSIC100 Flare §6.7 "Sealing installation instruction").
5 Completely loosen the pipe screw fitting.
6 Open the ball valve.
7 Push the sender/receiver unit into the pipeline.
8 Check the gasket of the pipe screw fitting for damage.
9 Screw in the pipe screw fitting and tighten with a torque of 150 Nm.
10 Push the sender/receiver unit further to the stop.
11 For probe version F1F-P, align the measuring path correctly before the self-cutting ring
fitting is fixed. For cross-duct versions, proceed with the next step.
12 Tighten the self-cutting ring fitting 1.25 turns.
Make sure the markings for the self-cutting ring fitting are next to each other again,
→ Fig. 32.
13 For cross-duct versions, install the Master sender/receiver unit on the nozzle on the
upstream side so that the sender/receiver unit points in the flow direction.
Observe when aligning the probe version
Align probe version F1F-P correctly before securing the self-cutting ring fitting again:
WARNING: Leakage hazard
Multiple use can damage the gasket of the pipe screw fitting.
▸
Before reuse, i.e. whenever the pipe screw fitting is to be tightened again,
check the gasket of the pipe screw fitting:
▸
Replace the gasket if it has visible deformations, indentations or damage. In
this case contact SICK Service.
Otherwise there is a risk of leakage.
WARNING: Danger due to gas leaks
When gas escapes, close the ball valve again and contact SICK Service.
WARNING: Leakage hazard
Multiple use can damage the gasket of the pipe screw fitting.
▸
Before reuse, i.e. whenever the pipe screw fitting is to be tightened again,
check the gasket of the pipe screw fitting:
▸
Replace the gasket if it has visible deformations, indentations or damage. In
this case contact SICK Service.
Otherwise there a risk of leakage.