27
8027872/1JTM/0-8/2023-05| SICK O P E R A T I N G I N S T R U C TI O N S | FLOWSIC550
Subject to change without notice
ELECTRICAL INSTALLATION 7
Safety data of Inputs and outputs
Terminal Function
active passive
Uo
[V]
Io
[mA]
Po
[mW]
Co
[µF]
Lo
[mH]
Ui
[V]
Ii
[mA]
Pi
[mW]
Ci
[µF]
Li
[mH]
Interfaces for interconnection with auxiliary equipment
DO0 Digital output 0
optical isolated
--- --- --- --- --- 20 --- 1100 0.024 ---
DO1 Digital output 1 --- --- --- --- --- 20 --- 1100 0.024 ---
DO2 Digital output 2 --- --- --- --- --- 20 --- 1100 0.024 ---
RS485 -1 RS485 Data interface,
external [Ex ia] power sup-
ply required
--- --- --- --- --- 15 --- 1100 IIA / D:
IIB /C, D:
IIC / A, B, C, D:
2.5
1.5
0.25
---
RS485 -2 RS485 Data interface,
external [Ex ia] power sup-
ply required
--- --- --- --- --- 15 --- 1100 IIA / D:
IIB / C, D:
IIC / A, B, C, D:
2.5
1.5
0.25
---
Interface for interconnection with auxiliary service equipment
M12
connector
Service/Bluetooth Dongle 8.2 410 668 IIA / D: 1000 µF
IIB / C, D: 81 µF
IIC / A, B, C, D: 7.6 µF
0.165 n/a n/a n/a n/a n/a