8027872/1JTM/0-8/2023-05| SICKO P E R A T I N G I N S T R U C T I O N S | FLOWSIC550
Subject to change without notice
8 COMMISSIONING
3 Press the push button to activate the service interface.
Fig. 8: Push button
4 Click on the icon to start FLOWgate
TM:
5 In FLOWgate
TM
click on icon “Scan” and scan for the device:
6 When the device has been found, click on “Connect”.
After successful login the Overview page is displayed.
Note:
The display and the service interface have a timeout and switch off after approx. 5 min-
utes (default setting) when no button has been pressed and no data transmission takes
place.
Default configuration of the service interface:
Protocol type: MODBUS-RTU
Baud rate: 38400
Protocol bits: 8N1