ACUSON Freestyle User Manual 6-7
Wireless Probes
Monitoring Wireless Signal Quality
The Freestyle has two methods of monitoring wireless signals:
Wireless Signal Quality meter
Average Noise Level indicator
Each is described in the following sections.
Wireless Signal Quality Meter
The signal meter has a total of six bars.
No Signal
OK
P S
No Signal
OK
1 2 3 4 5
Wireless Signal Quality Meter; showing ve different conditions
Number of Bars in the Wireless Signal
The number of bars in the wireless signal quality meter correlates to the potential of artifact
being introduced because of a weak signal.
Always monitor the number of bars displayed in the meter. See the following table for a
description of the conditions.
Wireless Signal Quality Meter Conditions
Number of Bars Condition Expectations
4 to 6 The best signal quality. Expect a very reliable wireless probe performance without
noise being introduced because of wireless link quality
3 Very reliable wireless probe performance, this noise is typically imperceptible, but
there is a possibility of low levels of noise (artifacts) being introduced into the image
because of a compromised wireless probe link.
Reorient the probe/console to within 3 meters of each other, in a line-of-sight,
without obstructions between the probe and the console.
2 or 1 There is an increased possibility of noise (artifacts) being introduced to the image
because of a compromised wireless probe link. These artifacts may appear as
bright, streaky lines. See the image in “B-Mode Wireless Noise” on page 6-10.
Reorient the probe/console to within 3 meters of each other, in a line-of-sight,
without obstructions between the probe and the console.
0 The wireless signal has been lost, so the image screen is blank. A message
appears on the console screen informing you of this condition.
Reorient the probe/console to within 3 meters of each other, in a line-of-sight,
without obstructions between the probe and the console.