8-10 ACUSON Freestyle User Manual
Connectivity
Conguring the System DICOM Settings
To congure the ACUSON Freestyle System DICOM settings:
1. Go to SETUP / SETTINGS / DICOM / SYSTEM.
2. Enter the system information as described in the following table.
DICOM System Settings Fields
Field Description
Freestyle AE Title Identies the AE Title assigned to the system.
Limited to 16 characters; cannot include the period (“.”) character.
Freestyle Port
Number
The port on the system used for DICOM messages.
The default is port 104.