ACUSON Freestyle User Manual 8-9
Connectivity
Conguring a Modality Worklist Server
To congure a DICOM Modality Worklist Server:
1. Go to SETUP / SETTINGS / DICOM / WORKLIST.
2. Select NEW.
3. Enter the DICOM Worklist information as described in the following table.
DICOM Worklist Fields
Field Description
Alias Name you dene for the server.
Limited to 16 characters.
IP Address The IP address of the Server.
AE Title Identies the server application that receives worklist messages from
the ACUSON Freestyle. The alphanumeric characters entered in this
eld must exactly match (including case) the Application Entity name
assigned to the server application.
Limited to 16 characters; cannot include the period (“.”) character.
Port Number The remote port on the server that is to receive the worklist messages.
The default is port 104.
Test Used to test the connection once you have entered the conguration
data. You can test using either Ping or C-Echo (DICOM Verify). Ping
tests the connection across the network; C-Echo additionally tests that
the DICOM server is active.
For security reasons, some network and network servers are set up
not to respond to Ping or C-Echo commands. Check with your IT
department or PACS Administrator to determine how your network
handles these functions.
NOTE: If you enable Worklist but your DICOM server does not support this feature,
the system will not perform the Worklist function.