ACUSON Freestyle User Manual 8-7
Connectivity
Conguring a DICOM Storage Server
To congure a DICOM storage server,
1. Go to SETUP / SETTINGS / DICOM / PACS.
2. Select New.
3. Enter the DICOM storage server information as described in the following table.
DICOM Storage Server Fields
DICOM Storage Server
Field
Description
Alias
A name you create to identify the DICOM server.
Limited to 16 characters.
IP Address The IP (Internet Protocol) address for the server. See your System
Administrator to determine your IP Address.
AE Title Identies the server application that receives the images from the
ACUSON Freestyle. The alphanumeric characters entered in this
eld must exactly match (including case) the Application Entity name
assigned to the server application.
Limited to 16 characters; cannot include the period (“.”) character.
Port Number The remote port on the server that is to receive the connection
request and the patient studies.
The default setting is 104.
Test Used to test the connection once you have entered the conguration
data. You can test using either Ping or C-Echo (DICOM Verify). Ping
tests the connection across the network; C-Echo additionally tests
that the DICOM server is active.
For security reasons, some network and network servers are set up
not to respond to Ping or C-Echo commands. Check with your IT
department or PACS Administrator to determine how your network
handles these functions.