EasyManua.ls Logo

Siemens ACUSON Freestyle - Page 61

Siemens ACUSON Freestyle
138 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ACUSON Freestyle User Manual 2-41
Safety and Care
WARNING: The polarity of the battery terminals is xed and cannot be switched or
reversed.
WARNING: Do not force the battery into its compartment in the probe.
WARNING: Do not connect the battery to an electrical power outlet.
WARNING: Do not continue to recharge a battery if it does not recharge properly after
two successive six hour charging cycles.
WARNING: If the battery leaks or emits an odor, remove it from all possible ammable
sources.
WARNING: Do not attempt to recharge non-rechargeable batteries, such as the batteries
included on printed circuit boards. Recycle non-rechargeable batteries according to local,
state, and regional regulations.
CAUTION: Remove the battery from the probe if the probe is not likely to be used for an
extended period. Prior to storage charge or discharge the battery to approximately 50% of
capacity.
CAUTION: To avoid the battery bursting, igniting, or emitting fumes from the battery and
causing equipment damage, observe the following precautions
CAUTION: Do not immerse the battery in liquid if it is damaged or the housing is cracked.
CAUTION: Do not immerse the battery in any liquid not listed in the chapter “Cleaning,
Disinfection, and Sterilization,” and do not submerge a battery beyond the time periods
specied in the instructions for recommended liquids.
CAUTION: If the battery emits an odor or heat, is deformed or discolored, or in any way
appears abnormal, immediately remove it from use.
CAUTION: Do not put the battery into a microwave oven or pressurized container.
CAUTION: Store the battery between 5º C and 60º C (41º and 140º F).
CAUTION: Use only Siemens-supplied batteries and charging devices.