9-2
9
Cleaning and Maintenance
3 Dry the test table thoroughly
(except for the white calibration
bar) with a soft cloth or lint-free
tissue.
4 Reinsert the test table into the
analyzer by holding the table at
the end opposite the white
calibration bar, with the white
calibration bar facing upwards.
Push the test table firmly but
slowly, just over halfway into
the analyzer.
5 Replace the test table insert.
Cleaning
Care should be taken not to
scratch the white calibration bar.
Instructions for cleaning the white
calibration bar are given later in this
section.
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.