Appendix G: Sample Interference Notes
Appendix G: Sample Interference Notes
Sample Interference Notes inform the user about test results that can be
affected by components detected in the urine sample.
Depending upon the strip and sample, Sample Interference Notes include
the following:
High SG may cause falsely lowered GLU results.
Elevated GLU may cause falsely lowered LEU results.
Visibly bloody urine may cause falsely elevated PRO results.
High SG may cause falsely lowered LEU results.
High pH may cause falsely elevated PRO results.