Testing
4-9
Make sure the test table insert is in
position for a cassette test.
Remove the test cassette from the
foil package and place the cassette
on the test table.
Touch START button.
The next screen that appears is
another Prepare Test. This screen
prompts you through the steps to
prepare the cassette test.
A timer displays how much time
you have remaining to complete the
steps.
Once you touch the START
button you have 8 seconds to
draw the urine sample into the
pipette and add the urine sample
into the well on the cassette.
Quick Tests