5-25
5
Instrument Set Up
Instrument Settings
Edit Reported Chemistries
This section describes how to
include or exclude urinalysis tests
from the reported results. Tests
include GLU, BIL, KET, SG, BLO,
pH, PRO, URO, NIT, LEU, ALB,
and CRE. The default is all tests
are reported and the button is
filled.
To set up tests for urinalysis,
perform the following steps:
1. At the Select Ready screen,
touch Instrument Set Up.
The Choose Settings screen
displays.
2. Use the arrow keys to select
Instrument Settings.
3. Touch Select.
The Instrument Settings screen
displays.
4. Use the arrow keys to select
Urinalysis Test Settings.
5. Touch Select.
The Urinalysis Test screen
displays.
6. Touch Edit reported
chemistries.
The Reported Chemistries
screen 1 of 2 displays.
7. To remove a selected test from
reported results, touch the
button for that test.
To include a test in reported
results, touch that test’s button
again.
8. Touch Next to advance to the
next screen.
The Reported Chemistries
screen 2 of 2 displays.
9. Touch the button for the test
you want to remove.
Touch the button again to
include that test.
10. Touch Done.
11.Touch Next.
12. Touch Done 3 times to return
to the Select Ready screen.
The reported results selection
applies to all strip types.