The red LED (fault) at the PPU is flashing (for version 1) or
flashing rapidly (for version 2)?
No analyzable signal is available.
◆ Ensure that the pulse sensor is attached correctly.
✓
The pulse sensor is attached.
◆ Perform the examination. See: Software operator manual
Respiratory triggering
Description
To keep respiratory artifacts to a minimum, respiratory triggering is
primarily used in abdominal imaging. Data acquisition for respiratory
triggering begins when the respiratory signal reaches a predefined
level (approx. 20 % of the maximum value). Respiratory movement is
minimal in this range.
(1) Expiration
(2) Inspiration
The respiratory signal is acquired using a respiratory cushion
connected via a pressure hose to the ECG and respiratory sensor
(PERU). The respiratory cushion is attached to the patient via the
respiratory belt.
Retrospective gating cannot be applied.
Performing
1 Ask the patient to lie still during the measurement.
2 Inform the patient that the knocking sounds during the
measurement are caused by switching the gradients on and off.
The PERU may also vibrate slightly.
Performing the examination
4.5
4.5.1
4.5.2
Informing the patient
4 Physiological imaging
100
MR System | Operator Manual
Print No. MR-01501G.621.02.02.02