The SYSTEM ON LED lights up. The MR system is switched on.
The software automatically starts at the syngo Acquisition Workplace.
Since the syngo MR Workplace has its own voltage supply, it is
switched on separately from the syngo Acquisition Workplace.
◆ Press the Power On switch at the computer of the syngo MR
Workplace.
The software of the syngo MR Workplace starts.
MAGNETOM Verio: After switching on the syngo Acquisition
Workplace and the syngo MR Workplace, the system requires
approx. 18 minutes to warm up and get ready for acquisition.
1 If a dialog box is displayed at the syngo Acquisition Workplace
informing you that the helium fill level is too low: Close the dialog
box and notify Siemens Service or have the magnet refilled.
2 Check all Table Stop buttons (at the intercom and at the control
units). Ensure that these buttons function properly and stop the
table immediately.
3 Check if pressing the squeeze bulb triggers the patient alert.
4 Check if communication with the patient in the examination room
works properly.
5 Check if image transmission of the video systems works properly.
6 Check if the contact spring connectors at the door frame and the
door to the examination room are free of residues, such as
cleaning agents, oil, grease, paint splatters, blood drops, etc.
Shutting down the system (System Off)
Shutting down the system includes the following steps:
◾ Shutting down the computer system
◾ Switching off the MR system at the alarm box
Switching on the
syngo MR
Workplace
Checking the MR system
components