EasyManua.ls Logo

Siemens MAGNETOM Verio - Performing

Siemens MAGNETOM Verio
140 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performing
The quality of the ECG signal for triggered measurements is enhanced
by:
Proper placement of the electrodes
Good skin contact of the electrodes
Reduction of interference signals caused by electromagnetic
induction.
Due to cable loops
Interferences from electrical potentials caused by muscle
movement
Attach the electrodes so that interferences from electrical
potentials caused by muscle movement and baseline drifts are
minimized. Suitable contact points are therefore areas that show
very little muscle or fatty tissue.
1 Prepare the patient for the examination as early as outside the
examination room.
2 On the chest of the patient, select locations with minimal muscle
and fat tissue for attaching the electrodes.
3 In case of hairy skin: shave the points where you intend to attach
the electrodes.
Shave the patient outside the examination room to prevent
accidents.
4 Thoroughly clean the patient's skin at the locations involved.
However, do not use solutions containing alcohol.
4.3.2
Image quality
Preparing the patient
Physiological imaging 4
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
97
Print No. MR-01501G.621.02.02.02

Table of Contents

Related product manuals