MR measurements with the footswitch are only possible for protocols
that were configured for manual start-up. This is the case e.g. for
protocols with MR measurements after administering contrast
medium.
If the protocol is configured for repeated measurements, the next
measurement can be started with the footswitch after completing the
preceding measurement. For more information regarding the
configuration of protocols, see: Software operator manual
1 Push the hose plug into the retaining rings of the pushbutton unit.
2 Load a suitable protocol and start it at the syngo Acq WP or the In-
Room syngo Acq WP.
The MR system is waiting for the manual start of the protocol.
3 Press the Start footswitch to begin the measurement.
As an alternative, you can start and end the measurement via the
In-Room syngo Acq WP or the syngo Acq WP.
4 Press the Stop footswitch to end the measurement.
Other components and accessories
Gradient supervision
To prevent damage to the MR system by a malfunction of the
gradient system, a specially designed supervision is installed at your
system.
This supervision monitors that cables, connections, or other
components of the gradient system do not show excessive heating. In
case of a malfunction, the measurement is stopped and an alarm
message is issued.
After 1 minute the system will be automatically switched to
Standby.
Performing the measurement
3.9
3.9.1
MR system components 3
MAGNETOM Avanto, MAGNETOM Verio | syngo MR E11
79
Print No. MR-01501G.621.02.02.02