Opdimar Register 3 SPB7-230.033.12 Page 18 of 26 Siemens-Elema AB
Installation Rev. 01 09.02 SPS-UD Solna, Sweden
5 - 18 Installation
4. Place the workstation table outside the patient environment of safety reasons.
Patient environment is determined by IEC 60601-1-1 as 1.5 m away from the
patient.
There must be enough space around the workstation table so that
it safely can be vertically adjusted.
150
Min.1,5 m
Cables 8 m
Cable 1,5 m
Biopsy
controller
CCD camera
Biopsy unit
Generator
Workstation
Cable duct
Stand
CCD camera holder
Cable (workstation table) 2 m
OPD00489
Fig. 19 System site overview
NOTE