EasyManua.ls Logo

Siemens MAMMOMAT 300 - Rotating anode cable X9

Siemens MAMMOMAT 300
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens-Elema AB SPS-UD RX B7-120.033.01.10.02 MAMMOMAT 300
Solna, Sweden 05.98 65 02 186
Cable connections 5-7
Rotating anode cable X9
Figure 7 Connecting X9 to the generator
1 Connect the rotating anode cable X9 to D711 X9 (1).
2 Connect the two blue free wires from connector D711 X9 (1) to capacitor C51 (2).
3 Fit the cable to the chassis by using a cable clamp (3).
CAUTION!
When fitting the cable clamp, ensure satisfactory contact between shield braid (4) and
chassis.
2
1
MAM00163
4
3

Table of Contents

Other manuals for Siemens MAMMOMAT 300

Related product manuals