Siemens-Elema AB SPS-UD RX B7-120.033.01.10.02 MAMMOMAT 300
Solna, Sweden 05.98 65 02 186
Checks without high voltage 10-1
Checks without high voltage
1 System "OFF".
2 On p.c. board D702 set switch S1 (UZW) to "ON" (upper position) and switch S2
(SS) to "OFF" (lower position).
3 System "ON".
4 Press the exposure buttons on the control panel. The rotating anode shall start up. At
the end of the limit time, the rotating anode is braked.
5 Release the exposure buttons.
NOTE!
If the exposure buttons are prematurely released, "Limit" will be indicated.
6 System "OFF".