Siemens MAMMOMAT 300 - Checks without high voltage

Siemens MAMMOMAT 300
93 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Siemens-Elema AB SPS-UD RX B7-120.033.01.10.02 MAMMOMAT 300
Solna, Sweden 05.98 65 02 186
Checks without high voltage 10-1
Checks without high voltage
1 System "OFF".
2 On p.c. board D702 set switch S1 (UZW) to "ON" (upper position) and switch S2
(SS) to "OFF" (lower position).
3 System "ON".
4 Press the exposure buttons on the control panel. The rotating anode shall start up. At
the end of the limit time, the rotating anode is braked.
5 Release the exposure buttons.
NOTE!
If the exposure buttons are prematurely released, "Limit" will be indicated.
6 System "OFF".