Siemens-Elema AB SPS-UD RX B7-120.033.01.10.02 MAMMOMAT 300
Solna, Sweden 05.98 65 02 186
Prerequisites 1-1
Meters and appliances, tools, notes
Meters and appliances required
NOTE!
Calibrated instruments are required.
Line resistance meter 44 15 733 RV090.
Protective ground-wire tester 44 15 899 RV090.
Oscilloscope, e.g TEKTRONIX 314.
Digital multimeter incl. mAs meter.
Densitometer e.g PDA 81.
Service PC (e.g Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable
(PC-Generator), part no. 96 00 440 RE999.
Normi 7 test body (provided by the customer) or SIB phantom Type 42 001 (PTW or
INAK).
Centring cross, part no. 96 60 051 RE 999.
Plexiglass plates, three plates measuring150 mm x 150 mm x 20 mm and one plate
measuring 150 mm x 150 mm x 10 mm, part no. 85 49 438 and 97 88 423 respective-
ly.
WARNING!
For safety reasons, the existing protective ground conductor in the power cord must un-
der no circumstances be disconnected when operating the oscilloscope. For those mea-
surements, in which any resulting ground loop may falsify the measuring result, use the
differential amplifier (difference measurement).
Tools required
Standard installation tool kit.
Torque wrench for bolting the stand/generator to the floor (optional).
Electrical screwdriver with adjustable torque is recommended.
NOTE!
Checks and /or adjustments, which must be made with X-ray radiation switched on, are
marked with the radiation warning symbol: