Siemens-Elema AB SPS-UD RX B7-120.033.01.10.02 MAMMOMAT 300
Solna, Sweden 05.98 65 02 186
Appendix 21-3
Measuring protocol
Values recorded during installation of Mammomat 300:
Keep measuring protocol in technical documentation.
Tube P49 Mo, molybdenum anode 0.3
Calculated values Required values
______V x 5 kV/V = _____kV 30 kV +/- 5 %
______V x ____ s x 40 mA/V = ____ mAs 20 mAs +/- 10 %
Date:______________
Sign:______________
Tube P49 Mo, molybdenum anode 0.15
Calculated values Required values
______V x 5 kV/V = _____kV 30 kV +/- 5 %
______V x ____ s x 40 mA/V = ____ mAs 10 mAs +/- 10 %
Date:______________
Sign:____________