Siemens-Elema AB SPS-UD RX B7-120.033.01.10.02 MAMMOMAT 300
Solna, Sweden 05.98 65 02 186
Options 20-1
Bolting the stand/generator to the floor
Should a permanent fastening to the floor be required due to local regulations (for exam-
ple in earthquake areas) or due to the quality of the floor covering, the stand/generator can
be secured to the floor by means of the levelling screws (1) of the stand or the fixing holes
(2) of the generator.
Figure 1 Base plate for the stand
Figure 2 Base plate for the generator
1 Position the stand/generator and mark out appropriate fixing holes on the floor, using
the base plate as a template.
NOTE!
The base plates shall be secured with at least four screws each (two on each side).
2 Drill the holes and insert dowels and screws use Liebig safety dowel M6 S10/35 and
washer DIN 9021B, ISO 7093 or equal.Tighten the screws to 10 Nm.
NOTE!
Make sure that the screws don’t come in contact with any conductive material such as re-
inforcing iron.
MAM00138
1
MAM00139
2