Siemens-Elema AB SPS-UD RX B7-120.033.01.10.02 MAMMOMAT 300
Solna, Sweden 05.98 65 02 186
Checking the compression device and OPCOMP 16-3
Checking the OPCOMP (Option)
See Figure 1
1 Decompress.
2 Set the compression force to 20 kg by using the presetting knob (5) and compress.
The compression movement must be interrupted by OPCOMP before 20 kg is
reached, i.e. at 5-15 kg depending on the material (4) used.