Page 104 Milltronics SF 500 - INSTRUCTION MANUAL 7ML19985CN01
mmmmm
Parameters
P358 On-line Calibration Activation
Initiates on-line calibration.
Entry:
0 = OFF
f
1 = ON
P359 Factoring
Factoring is used as a method of calculating the value of the test rate (P017) to a new test
reference. The task is performed only for the weight relevant for the multispan selected, if
applicable.
Entry:
1 = weight (
f
= 1)
Refer to
Recalibration
on page 30 for execution of the factoring procedure.
P360 Calibration Duration
Sets the number of whole calibration durations used during a zero or span calibration.
(
f
= 1 which is approximately 20 seconds.)
Enter number of time periods, range 1 to – 99.
P365 Multispan
Select the span reference to be applied for determination of rate and totalization.
Entry:
1 = multispan 1 (MS1), for product or condition A
f
2 = multispan 2 (MS2), for product or condition B
3 = multispan 3 (MS3), for product or condition C
4 = multispan 4 (MS4), for product or condition D
5 = multispan 5 (MS5), for product or condition E
6 = multispan 6 (MS6), for product or condition F
7 = multispan 7 (MS7), for product or condition G
8 = multispan 8 (MS8), for product or condition H
Refer to
Multispan
on page 38 and
P270 Auxiliary Input Function
(6) on page 100.
Note: Totalization is halted during the factoring procedure, and resumed
only upon return to the RUN mode.