Siemens Milltronics SF 500 - User Manual

Siemens Milltronics SF 500
134 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Instruction Manual September 2003
SF 500
milltronics

Table of Contents

Questions and Answers

  • P
    Philip BlackAug 18, 2025
    What to do if I get a 'Remote Totalizer 1 or 2 Exceeded' error message on Siemens Milltronics SF 500?
    • A
      Anthony WallaceAug 18, 2025
      This error indicates the remote totalizer buffer has overflowed. You can resolve this by increasing the remote totalizer resolution or clearing the internal totalizer reset (P648).
  • C
    christophermartinezAug 20, 2025
    What causes Siemens Measuring Instruments to display 'BF'?
    • K
      Kathryn WhiteAug 20, 2025
      The 'BF' on the display indicates a Battery Failure. Check to see if the battery needs to be installed, or replaced.
  • L
    Lee CervantesAug 23, 2025
    What does it mean when an error message is flashing on the Siemens Measuring Instruments display?
    • R
      Richard CarrAug 23, 2025
      An error message flashing on the display may indicate an error that occurred some time ago. Verify if the error is still active by cycling the SF 500 from PROGRAM to RUN modes.
  • S
    Sheena WoodAug 25, 2025
    Why is the display blank on my Siemens Measuring Instruments?
    • M
      Micheal ThomasAug 26, 2025
      A blank display on your Siemens Measuring Instruments can stem from several causes: * Improper power source wiring: Check the wiring. * Incorrect supply voltage: Ensure the proper voltage is applied. * Power selection switch in the wrong position: Verify its correct placement. * Faulty fuse: Check the fuse's condition and installation. * Loose cables: Confirm the cables between the display and motherboard are properly connected.
  • K
    kristen67Aug 28, 2025
    What to do if the Siemens Milltronics SF 500 Measuring Instruments keypad is not operative?
    • W
      wileymarissaAug 28, 2025
      If the keypad on your Siemens Measuring Instruments isn't working, consider these potential causes: * Improper cabling: Ensure the cable connecting the keypad to the motherboard is intact. * SW1 in the left position: Switch SW1 (underneath the display) to the right. * Incorrect security settings: Check the security code P000 and security level P350.
  • L
    Laura AyalaAug 30, 2025
    Why can't I print from my Siemens Milltronics SF 500 Measuring Instruments?
    • B
      breannapattersonAug 30, 2025
      If you're unable to print, it could be due to: 1. Missing programmed parameters: Ensure all necessary parameters are programmed. 2. Improper wiring: Check the wiring between the printer and SF 500. 3. Incorrect mode: Switch to RUN mode to enable printing.
  • J
    jessechambersSep 2, 2025
    How to fix poor communications using Modbus RTU and ASCII with Siemens Milltronics SF 500?
    • R
      Ruth McleanSep 2, 2025
      Poor communication using Modbus RTU and ASCII can be caused by improper wiring or shielding. Ensure proper wiring and shielding. Also, try a different Baud Rate or decrease the request length.
  • J
    Joseph YoungSep 4, 2025
    What to do if I changed a communications parameter on Siemens Measuring Instruments but it's not working?
    • K
      Kenneth JeffersonSep 4, 2025
      If a communications parameter change isn't working, cycle the power to the SF 500. The SF 500 power must be cycled prior to any communications parameter taking effect.
  • M
    Mary MooreSep 7, 2025
    Why does the Siemens Milltronics SF 500 Measuring Instruments rate display seem chaotic?
    • Z
      zsimsSep 7, 2025
      A chaotic rate display can be caused by improper wiring from the LVDT or load cells. Check the wiring. It may also be caused by incorrectly set millivoltage levels. Measure millivoltage levels across SF 500 terminals 1 & 2 and compare to P940 01 (Load Cell A) and, if applicable, across SF 500 terminals 3 & 4 and compare to P940-02 (Load Cell B). If there are major differences in readings, contact Siemens Milltronics.
  • D
    dalelopezSep 9, 2025
    What to do if the Siemens Milltronics device parameter remains unchanged?
    • L
      Linda ReedSep 9, 2025
      If a parameter remains unchanged, it could be due to an external error. Use the keypad to set the parameter. It could also be that the parameter is locked; check the lock parameter (P000). Finally, check if SW1 is in the left position and switch it to the right.

Summary

Installation

Start Up

Load Cell Balancing

Procedure for electronically balancing the load cells.

Zero Calibration

Process for setting the zero point of the weighing system.

Span Calibration

Process for setting the span or full-scale calibration.

Recalibration

Material Tests

Procedure to verify calibration accuracy using material samples.

Routine Zero

Steps for periodic recalibration of the zero point.

Routine Span

Steps for periodic recalibration of the span.

PID Control

Setpoint Controller - Rate Control

Configuration for controlling feed rate via setpoint.

Setup and Tuning

Procedures for tuning PID control parameters.

Parameters

Start Up (P001 to P017)

Essential parameters for initial device setup and configuration.

PID Control Parameters (P400 - P419)

Parameters for configuring PID control loops.

Batch Control (P560 - P568)

Parameters for managing batch operations and sequences.

Troubleshooting

Common Problems Chart

Lists common issues, their causes, and recommended solutions.

Siemens Milltronics SF 500 Specifications

General IconGeneral
TypeUltrasonic Flowmeter
Measurement Range0.1 to 12 m/s
Enclosure RatingIP65
Measurement PrincipleTransit Time Difference
Process Temperature-40 to 200 °C
Operating Temperature-20 to +60 °C (-4 to +140 °F)
Process PressureUp to 25 bar
Output Signal4-20 mA, HART
MaterialAluminum, Stainless Steel
Accuracy±0.5% of flow rate (for flow velocity > 0.3 m/s (1 ft/s))
Pipe Size Range25 mm to 1200 mm (1 inch to 48 inch)