R
Rachel DennisSep 12, 2025
How to resolve LVDT Conditioning Card malfunction in Siemens Milltronics SF 500 Measuring Instruments?
- CChristina WhiteSep 12, 2025
If the LVDT Conditioning Card doesn't function, check the interconnecting wiring between the LVDT and conditioning card, and between the conditioning card and the SF 500. Also, check to ensure that SF 500 terminals 2 and 17 are interconnected.