7ML19985CN01 Milltronics SF 500 - INSTRUCTION MANUAL Page 75
mmmmm
Communications
P781 Data message
Sets the data message to be delivered via the selected port, ports 1 to 3 (P781-01 to -03).
All messages and printouts include time and date.
Entry:
0 = no message
f
1 = rate
2 = total*
3= rate and total*
4 = batch total (totalizer 5)
5 = quick start parameters (P001 – P017)
6 = all parameters
*totalizer 1 and/or 2 as set by P647, Totalizer Display
P799 Communications Control
Assigns programming control either locally through the keypad or Dolphin Plus (P770 = 1),
or remotely through Modbus protocol (P770 = 2 or 3) or SmartLinx
®
.
Entry:
0 = local
1 = remote
Note: Applicable only to ports programmed for printer communication (parameter
770).
WARNING: When a SmartLinx
®
card is installed and P799 = 1, the
parameters that the SmarlLinx
®
card is writing to the SF 500 will
be continuously updated. Therefore, if you connect a SmartLinx
®
card to the SF 500, set P799 = 1 and not write anything to the
SmarLinx
®
card, your setpoints will be 0.