EasyManua.ls Logo

Siemens MOBILETT XP - Adjusting the Radiation Field

Siemens MOBILETT XP
109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
Operator Manual
SPR8-230.621.01.03.02
40 of 106
Adjusting the radiation field
Switch on the collimator light.
Set the appropriate film focus distance, using the built-in measuring tape.
Turn the collimator until the light field is parallel with the sides of the cassette.
Place the detachable protractor on the cassette.
Read the inclination angle and adjust it.
Collimate the radiation field for the desired image size.