EasyManua.ls Logo

Siemens MOBILETT XP - Radiation Calculation and Dose Estimation; Dose Area Product (DAP) Calculation

Siemens MOBILETT XP
109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Messages
MOBILETT
SPR8-230.621.01.03.02 57
of 106
Messages
System Messages
Acknowledging system messages
Press any button (kV, mAs, collimator).
The system message disappears. The MOBILETT is ready for operation.
The message remains or appears again.
Please notify Siemens Customer Service.
If the message “Pause” appears, the capacitor loader is overheating. Wait until
the message disappears.
If error 3 (ERR 3) appears: Switch the unit off and wait until the single tank cools
down. See the tube housing cooling curve.
If the message (ERR 25) appears, exposure has not been triggered within the
max. preparation time of 15 seconds.