Annual Maintenance
Maintenance every 12 months is necessary to keep the equipment safe and oper-
ational.
We recommend that the maintenance is performed by the Siemens customer
service.
Check of the system
Check of system parts
Interval Operations to perform
12 months Mechanical test
12 months Functional check
12 months Checking of image quality
12 months Checking of kV/mAs accuracy
12 months Electrical safety checks
12 months USA only:
Checking of reproducibility
Interval Material Material No.:
12 months Collimator lamp 8392016
24 months Battery pack (left) BK1/BK2 08392024
24 months Battery pack (right) BK3/BK4 08392032
48 months Cable winder complete 06508746
12 months Batteries remote control (Option) 1 x 9 V alkaline

Table of Contents