Radiation Calculation
Operator Manual
SPR8-230.621.01.03.02
64 of 106
Radiation Dose
The absorbed radiation dose (skin dose), D (Gy), can be estimated from the values
of Y in Diagram 1.
The absorbed radiation dose (skin dose) is dependent on:
the x-ray tube voltage, U (kV)
the quantity of charge, Q (mAs))
the source to skin distance, r (m)
the permanent (inherent) filtration of the single tank (mm Al)
Estimate the radiation dose
The radiation dose D is calculated with Equation (2):
Equation (2): D = Y x Q / r
2
Y (µGy x m
2
/mAs) - conversion factor
Q (mAs) - quantity of charge during exposure
r (m) - source to skin distance. Use the built-in measuring tape to determine r
Read the value of Y (µGy x m
2
/mAs) from Diagram 1, corresponding to the
tube voltage (kV) used during exposure.
Calculate the radiation dose D with Equation (2).