Radiation Calculation
MOBILETT
SPR8-230.621.01.03.02 63
of 106
Calculate the dose area product with Equation (1):
DAP = 57 (µGy x m
2
/mAs) x 10 x 10 x 10 (mAs) x 6.40 (cm
2
/m
2
)
= 364800 µGycm
2
Since 1000 µGy = 1 mGy, DAP = 364 800 µGycm
2
365 mGycm
2