Cleaning and Disinfecting
Operator Manual
SPR8-230.621.01.03.02
76 of 106
Disinfection
For surface disinfection, we recommend the use of isopropyl alcohol. Disinfec-
tants based on substituted phenols and compounds splitting of chlorine are
caustic and therefore, cannot be generally recommended.
As a general rule, no disinfectants sprays must be used, since the spray mist may
penetrate into the equipment which means the safety of the equipment can no
longer be assured (damage to electronic components, formation of flammable air
solvent vapour mixtures).
Use fluid disinfectants only.
As is known, some components of disinfectants are detrimental to health. Their
concentration in respiratory air must not exceed a legally determined limit. We
recommend that the appropriate directions on use issued by the manufacturers
of these agents be strictly observed.