The Exposure Point System
MOBILETT
SPR8-230.621.01.03.02 69
of 106
Example 3 Keeping the density the same, but changing contrast
By pressing the kV- button once and the mAs+ button twice, the density will
remain the same but yield higher contrast.
66 kV70 kV
20 mAs 25 mAs