Dosimetry
MOBILETT
SPR8-230.621.01.03.02 95
of 106
Pelvis
Upper extremities
Object Position SID KV mAs Grid Sensitivity
class
Pelvis/Hip a. 115 73 8 w.o. 400
Sacrum a. 115 73 12.5 w.o. 400
Sacrum lat. 115 90 20 w.o. 400
Object Position SID KV mAs Grid Sensitivity
class
Shoulder . a. 115 60 8 w.o. 200
Shoulder . axial 105 57 8 w.o. 200
Upper arm a./lat. 105 52 5 w.o. 200
Elbow a. 105 52 3.2 w.o. 200
Elbow lat. 105 52 4 w.o. 200
Forearm a. 105 52 3.2 w.o. 200
Forearm lat. 105 52 4 w.o. 200
Wrist d.v. 105 52 1.25
2.5
w.o. 200
100
Wrist lat. 105 52 2.5
5
w.o. 200
100
Hand d.v. 105 52 1.25
2.5
w.o. 200
100
Hand lat./obl. 105 52 1.6
3.2
w.o. 200
100
Finger 105 52 1.25
2.5
w.o. 200
100