ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
125
7.5.3 Transducer Block TB 1
Transducer Block TB 1 (Analyser)
Class
S
lo
t
Index
abs
Ind
ex
rel
Elem
ent
Default
V
alue
O
bject Parameter description Data Type Num
be
Store
(acc. to
profile)
cy
c
li
c
rea
d
w
rit
e
BLOCK
1 66 0 BLOCK_OBJECT Block characteristic
d
s32 1 C x
1
50 Reserved 250 = not used
2
3
Block Object Block-Typ (Function)
3
7
Parent Class Parent class: 01 = Input
41
C
lass 01 = Analog Input
5
0
DD-Reference reserved
6
0
DD-Revision reserved
7
0
x4002 Profile PA-profil-number assigned by PNO for
PROFIBUS - PA = 64
8
0
x0300 Profile Revision Profile 3.0
9
0
Execution Time f0r future use
10
3
6 NumberOfParameters number of parameters
11
0
x016d
A
ddress of VIEW 1 Index of view 1
12 1 Number of View Lists Number of views
S
TD 67 1
0
S
T_REV Identification / device / statistic rev.-no. /
Function Block: counter which is incremented
at any change of configuration parameters
unsigned16 1 N x
S
TD 68 2
T
AG_DESC Unique tag in the system which can be used
by the user
v
isible_string 32 S x x
S
TD 69 3
0
S
TRATEG
Y
Strategy unsigned16 1 S x x
S
TD 70 4
0
A
LERT_KE
Y
User can write value for alert handling unsigned8 1 S x x
S
TD 71 5
0
x08
T
ARGET_MODE Target mode of block = Auto unsigned8 1 S x x
S
TD 72 6 MODE_BLK Set block mode
d
s37 1 D x
1
0
x08
A
ctual Actual mode
2
0
x08 Permitted Permitted modes: Auto, O/S, MAN
3
0
x08 Normal Set mode in normal operation
S
TD 73 7
A
LARM_SUM Alarm status
d
s42 1 D x
1
0
x0000
C
urrent_Alarm Current Alarm
2
0
x0000 Unacknowledged Not acknowledged
3
0
x0000 Unreported Not reported
4
0
x0000 Disabled Disabled
MAND 74 8
C
omponen
t
1
C
OMPONENT_NAME Name of the measurand / component
v
isible_string 32 S x x
MAND 75 9 PV Measured value for AI-Block 1
d
s60 1 D x
0
.0
V
alue
0
x0B
S
tatus
T
ime
MAND 76 10 1997 PV_UNIT Unit of Primary Value (1997 = none) unsigned16 1 S x x
MAND 77 11 PV_UNIT_TEXT Unit Text of Primary Value
v
isible_string 8 S x x
MAND 78 12 1
A
CTIVE_RANGE unsigned8 1 S x x
MAND 79 13
0
A
UTORANGE_ON boolean 1 S x x
MAND 80 14
00
S
AMPLING_RATE
s
ampling rate in milliseconds unsigned32 1 S x x
81 15 bis
r
eserved byPNO
90 24
MAND 91 25 1 NUMBER_OF_RANGES unsigned8 1 C x
MAND 92 26 RANGE_1
d
s61 1 N x x
0
.0 Begin
100.0 End
V
IEW 109 43
V
IEW_1
View-Objekt 1
v
oid 5 D x
1ST_REV
6
MODE_BLK
7
PV
9
ALARM_SUM
12 ACTIVE_RANGE