ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
85
2.6.3 ULTRAMAT 6 F / OXYMAT 6 F / CALOMAT 6 F Pin Assignment
You will find a detailed description in the respective device manual (chapter 2).
Terminal block
Function in
device
PIN Function Comments
M 1GND
2
3 Relay 7 Contact loading:
4 max. 24 V / 1A, AC/DC
5
6 Relay 8
7 Relay is dead
8 for relay contact setting shown
9 Relay 9
10
11
12 Relay 10
13
14
15 Relay 11
16
17
18 Relay 12
19
20
21 Relay 13
22
23
24 Relay 14
25
26 Digital input 7...14 - N
27 Digital input 7 - P Potential-free
28 Digital input 8 - P via optocoupler
29 Digital input 9 - P
30 Digital input 10 - P "0" = 0 V ( 0...4,5V )
31 Digital input 11 - P "1" = 24 V ( 13...33V )
32 Digital input 12 - P
33 Digital input 13 - P
34 Digital input 14 - P
35 NC (not connected)
36 NC (not connected)
M 37 GND