ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
130
7.5.6 Transducer Block 2
Transducer Block TB2 (Analyser)
Class Sl
ot
Index
abs
Ind
ex
rel
Elem
ent
Default
Value
Object Parameter description Data Type Num
ber
Store
(acc. to
profile)
cy
cli
c
re
ad
wri
te
BLOCK
2 66 0 BLOCK_OBJECT Block characteristic ds32 1 C x
1 250 Reserved 250 = not used
2 3 Block Object Block-Typ (Function)
3 7 Parent Class Parent class: 01 = Input
4 1 Class 01 = Analog Input
5 0 DD-Reference reserved
6 0 DD-Revision reserved
7 0x4002 Profile PA-profil-number assigned by PNO for
PROFIBUS - PA = 64
8 0x0300 Profile Revision Profile 3.0
9 0 Execution Time f0r future use
10 27 NumberOfParameters number of parameters
11 0x026d Address of VIEW 1 Index of view 1
12 1 Number of View Lists Number of views
STD 67 1 0 ST_REV Identification / device / statistic rev.-no. /
Function Block: counter which is incremented
at any change of configuration parameters
unsigned16 1 N x
STD 68 2 TAG_DESC Unique tag in the system which can be used
by the user
visible_string 32 S x x
STD 69 3 0 STRATEGY Strategy unsigned16 1 S x x
STD 70 4 0 ALERT_KEY User can write value for alert handling unsigned8 1 S x x
STD 71 5 0x08 TARGET_MODE Target mode of block = Auto unsigned8 1 S x x
STD 72 6 MODE_BLK Set block mode ds37 1 D x
1 0x08 Actual Actual mode
2 0x08 Permitted Permitted modes: Auto, O/S, MAN
3 0x08 Normal Set mode in normal operation
STD 73 7 ALARM_SUM Alarm status ds42 1 D x
1 0x0000 Current_Alarm Current Alarm
2 0x0000 Unacknowledged Not acknowledged
3 0x0000 Unreported Not reported
4 0x0000 Disabled Disabled
MAND 74 8 Compone
nt 2
COMPONENT_NAME Name of the measurand / component visible_string 32 S x x
MAND 75 9 PV Measured value for AI-Block 1 ds60 1 D x
0.0 Value
0x0B Status
Time
MAND 76 10 1997 PV_UNIT Unit of Primary Value (1997 = none) unsigned16 1 S x x
MAND 77 11 PV_UNIT_TEXT Unit Text of Primary Value visible_string 8 S x x
MAND 78 12 1 ACTIVE_RANGE unsigned8 1 S x x
MAND 79 13 0 AUTORANGE_ON boolean 1 S x x
MAND 80 14 400 SAMPLING_RATE
s
ampling rate in milliseconds unsigned32 1 S x x
81 15 bis reserved by PNO
90 24
MAND 91 25 1 NUMBER_OF_RANGES unsigned8 1 C x
MAND 92 26 RANGE_1 ds61 1 N x x
0.0 Begin
100.0 End
VIEW 109 43
VIEW_1
View-Objekt 1 void 5 D x
1ST_REV
6MODE_BLK
9
PV
7 ALARM_SUM
12 ACTIVE_RANGE