EasyManua.ls Logo

Siemens OXYMAT 6 - Physical Block PB

Siemens OXYMAT 6
140 pages
Go to English
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
60
7.5.4 Physical Block
Physical Block PB
Class Sl
ot
Index
abs
Ind
ex
rel
Elem
ent
Default
Value
Object Parameterbeschreibung Data Type Num
ber
Store
(acc. to
profile)
cy
cli
c
re
ad
wri
te
BLOCK
1 110 0 BLOCK_OBJECT Block-Characteristik ds32 1 C x
1 250 Reserved 250 = not used unsigned8 1
2 1 Block Object Block-Typ (Physical) unsigned8 1
3 5 Parent Class 5 = Analyser unsigned8 1
4 250 Class 250 = not used unsigned8 1
5 0 DD-Reference reserved unsigned32 1
6 0 DD-Revision reserved unsigned16 1
7 0x4002 Profile PA-Profil-Nummer durch PNO vergeben für
PROFIBUS - PA = 64, Class B=2
unsigned16 1
8 0x0300 Profile Revision Profil 3.0 unsigned16 1
9 0 Execution Time for future use unsigned8 1
10 54 NumberOfParameters High Byte:Highest Index -Low Byte: Highest
Offset
unsigned16 1
11 0x01b7 Address of VIEW 1 unsigned16 1
12 1 Number of View Lists unsigned8 1
STD 111 1 0 ST_REV Zähler,der bei jeder Änderung von
Konfigurationsparametern inkrementiert wird
unsigned16 1 N x
STD 112 2 Analyser TAG_DESC eindeutige Tag im System, die der Anwender
spezifizieren kann
visible_string 32 S x x
STD 113 3 0 STRATEGY Strategy unsigned16 1 S x x
STD 114 4 0 ALERT_KEY Wert kann vom Anwender für die
Alarmbehandlung geschrieben werden
unsigned8 1 S x x
STD 115 5 0x08 TARGET_MODE Zielmodus des PB = Auto unsigned8 1 S x x
STD 116 6 MODE_BLK eingestellter Blockmodus ds37 1 D x
1 0x08 Actual Auto
2 0x88 Permitted Auto
3 0x08 Normal Auto
STD 117 7 ALARM_SUM Alarmstatus ds42 1 D x
1 0x0000 Current_Alarm
2 0x0000 Unacknowledged
3 0x0000 Unreported
4 0x0000 Disabled
MAND 118 8 /
0.23
SOFTWARE_REVISION Software Revision visible_string 16 D x
MAND 119 9 HARDWARE_REVISION Hardware Revision visible_string 16 D x
MAND 120 10 42 DEVICE_MAN_ID Hersteller unsigned16 1 C x
MAND 121 11 DEVICE_ID Geräteidentifikation (unique) visible_string 16 C x
MAND 122 12 DEV_SER_NUM Geräteseriennummer visible_string 16 D x
MAND 123 13 0x000000
00
DIAGNOSIS Detaillierte Geräte-Info(bitweise codiert) octet_string 4 D x
OPT 124 14 DIAGNOSIS_EXTENSION Erweiterte Diagnose octet_string 6 D
MAND 125 15 0xBBBF0
000
DIAGNOSIS_MASK Bits supported by DIAGNOSIS octet_string 4 C x
OPT 126 16 DIAGNOSIS_MASK_EXTENSION Bits supported by DIAGNOSIS_EXTENSION octet_string 6 C
OPT 127 17 CE,
NE21. Ex-
certificatio
n see
plate
DEVICE_CERTIFICATION erfüllte Zertfiziereungen Ex, CE visible_string 32 C x
OPT 128 18 2457 WRITE_LOCKING Password unsigned16 1 N x x
OPT 129 19 0 FACTORY_RESET Rücksetzen auf Default-Werte unsigned16 1 S x x
OPT 130 20 descriptor DESCRIPTOR visible_string 32 S x x
OPT 131 21 message DEVICE_MESSAGE Nachricht visible_string 32 S x x
OPT 132 22 01.01.200
0
DEVICE_INSTAL_DATE Datum der Geräteinstallation octet_string 16 S x x
OPT 133 23 0 LOCAL_OP_ENA Freigabe Vor Ort Bedienung unsigned8 1 N x x
MAND 134 24 1 IDENT_NUMBER_SELECTOR Profile Analyser spez. (0) Profile spec. (3)
Hersteller spez. (1) DP-Ident Number
unsigned8 1 S x x
OPT 135 25 HW_WRITE_PROTECTION Hardware-Schreibschutz unsigned8 1 D x
136 26 bis reserviert durch PNO
142 32