ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
91
3.1. Profiles for ULTRAMAT, OXYMAT and CALOMAT
Block model for PROFIBUS-PA
PROFIBUS PA Option.ULTRAMAT, OXYMAT
or CALOMAT
Sensor
Measured
value
processing
Function Block
Analog Input
PROFIBUS PA /
Physical Block
Device
Measured value =
input value
Out
ut value
Transducer
Block
Block structure of a gas analyzer with 4 block types, example for one measured value.
☛ Note
The complete table of objects you will find in appendix chapter 7.5
Transducer Blocks Function Blocks
Variants
TB AI
( Analog Input )
DO
( Digital Output )
ULTRAMAT 6 with 1 component 1 1 1
ULTRAMAT 6 with 2 components 2 2 1
ULTRAMAT 23 with 1 component 1 1 1
ULTRAMAT 23 with 2 components *) 2 2 1
ULTRAMAT 23 with 3 components *) 3 3 1
ULTRAMAT 23 with 4 components *) 4 4 1
OXYMAT 6 1 1 1
CALOMAT 6 1 1 1
OXYMAT 61 1 1 1
On ULTRAMAT 6 and ULTRAMAT 23 the assignments to the infrared (IR) measuring components always
start with “1“ in ascending order.