ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
80
2.4 PROFIBUS-PA Bus Connectors
The 9-pin Sub-D female connector enclosed with the device/option is required.
The bus cable can be connected with a solder joint.
Connectors with removable terminal connections are available, for example, from
Company: Phoenix Contact
Type: SUBCON 9/F-SH
Item no.: 27 61 49 9
2.4.1 Pin Assignment PROFIBUS-PA Bus Connector
Function PIN
PA-P ( + )
3
PA-N ( - )
7,8
NC (Not connected) 1,2,4,5,6,9
2.4.2 PROFIBUS-PA Cable
s - order number: (cables for FastConnect technique)
6XV1 830-5CH10 (blue) for applications in hazardous areas
6XV1 830-5DH10 (black) for applications in non-hazardous areas
Note
The cable screen should be connected fully in the connector housing.