3 - 1
Siemens AG RXB8-115.061.02 Page 1 of 2 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Error messages 3
There are three types of errors: 3
The following error codes are shown in the kV or mAs display on the control panel.
Initialization error
90... EPROM check sum error
95... IH
nom
circuit error
96... mA
nom
circuit error
97... kV
nom
circuit error
Stand-by error:
03... IH
act
< IHstby
04... IH
act
> IHstby
05... kV
act
< > 0
06... mA
act
< > 0
09... High pressure in the tank
Exposure error:
11... Inverter short circuit
12... kV
max
13... I
max
14... kV
act
< kVnom
15... mA
act
< mAnom
16... mA
act
> mAnom
17... maximum exposure time
18... exposure not completed