Notes on trouble-shooting 4 - 17
Siemens AG RXB8-115.061.02 Page 17 of 24 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Checking the kV and tube current (IR) 4
• Connect oscilloscope on D910 to MP "kV", "IR" and "GND".
• POLYMOBIL ON
• Trigger exposure with the default values 60 kV, 10 mAs.
The transient response of the tube current can be set with P1 on
D910.
kV
IR
Prolonged transient response
kV
IR
NOTICE