Replacing important components 5 - 3
Siemens AG RXB8-115.061.02 Page 3 of 6 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Replacing the light-beam localizer lamp 5
• The replacement light-beam localizer lamp is located below the cassette tray.
• Turn both knobs (8/Fig. 4) to an end position and remove them.
• Remove front panel after loosening both screws (7/Fig. 4) .
Caution: The frontpanel is connected to the collimator via cables (9/Fig. 5).
• Loosen both screws (10/Fig. 5) and unscrew the bracket carrier of the temperature
switch).
• Take out the defective lamp (11/Fig. 5) and replace it by a new lamp.
• Screw down the bracket and the front panel.
• Adjust both knobs and fasten them.
Do not touch the glass envelope with your bare fingers.
9
10
11
8
7
Fig. 4 Fig. 5
HINWEIS:
CAUTION: