Replacing important components 5 - 5
Siemens AG RXB8-115.061.02 Page 5 of 6 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Fig. 7
Replacing the caster 5
Herefore:
• Install the new caster in the reverse order.
• Check pedal positions (see operating instructions)
TD RX 6/ Friedrich
TD RX 1 / Kern R.
TDU 3 / Hay
1 Only loosen the Allen screws at the right and left (key width 2.5 mm)!
2 Move hexagonal shaft backwards
3 Remove pedal
4 Remove allen screws (key size 6 mm)
5 Remove caster.
6 Bring the threaded holes of the new caster into position